Vulnerabilities (CVE)

Filtered by vendor Medtronic Subscribe
Filtered by product Minimed 508
Total 1 CVE
CVE Vendors Products Updated CVSS v2 CVSS v3
CVE-2019-10964 1 Medtronic 38 Minimed 508, Minimed 508 Firmware, Minimed Paradigm 511 and 35 more 2024-11-21 5.8 MEDIUM 8.8 HIGH
In Medtronic MinMed 508 and Medtronic Minimed Paradigm Insulin Pumps, Versions, MiniMed 508 pump – All versions, MiniMed Paradigm 511 pump – All versions, MiniMed Paradigm 512/712 pumps – All versions, MiniMed Paradigm 712E pump–All versions, MiniMed Paradigm 515/715 pumps–All versions, MiniMed Paradigm 522/722 pumps – All versions,MiniMed Paradigm 522K/722K pumps – All versions, MiniMed Paradigm 523/723 pumps – Software versions 2.4A or lower, MiniMed Paradigm 523K/723K pumps – Software, versions 2.4A or lower, MiniMed Paradigm Veo 554/754 pumps – Software versions 2.6A or lower, MiniMed Paradigm Veo 554CM and 754CM models only – Software versions 2.7A or lower, the affected insulin pumps are designed to communicate using a wireless RF with other devices, such as blood glucose meters, glucose sensor transmitters, and CareLink USB devices. This wireless RF communication protocol does not properly implement authentication or authorization. An attacker with adjacent access to one of the affected insulin pump models can inject, replay, modify, and/or intercept data. This vulnerability could also allow attackers to change pump settings and control insulin delivery.