Filtered by vendor Medtronic
Subscribe
Filtered by product 2090 Carelink Programmer Firmware
Subscribe
Total
3 CVE
CVE | Vendors | Products | Updated | CVSS v2 | CVSS v3 |
---|---|---|---|---|---|
CVE-2018-5448 | 1 Medtronic | 2 2090 Carelink Programmer, 2090 Carelink Programmer Firmware | 2024-11-21 | 2.7 LOW | 5.7 MEDIUM |
All versions of the Medtronic 2090 Carelink Programmer are affected by a directory traversal vulnerability where the product's software deployment network could allow an attacker to read files on the system. | |||||
CVE-2018-5446 | 1 Medtronic | 2 2090 Carelink Programmer, 2090 Carelink Programmer Firmware | 2024-11-21 | 2.1 LOW | 5.3 MEDIUM |
All versions of the Medtronic 2090 Carelink Programmer are affected by a per-product username and password that is stored in a recoverable format which could allow an attacker with physical access to a 2090 Programmer to obtain per-product credentials to the software deployment network. | |||||
CVE-2018-10596 | 1 Medtronic | 2 2090 Carelink Programmer, 2090 Carelink Programmer Firmware | 2024-11-21 | 5.2 MEDIUM | 8.0 HIGH |
Medtronic 2090 CareLink Programmer all versions The affected product uses a virtual private network connection to securely download updates. The product does not verify it is still connected to this virtual private network before downloading updates. An attacker with local network access to the programmer could influence these communications. |