All versions of the Medtronic 2090 Carelink Programmer are affected by a per-product username and password that is stored in a recoverable format which could allow an attacker with physical access to a 2090 Programmer to obtain per-product credentials to the software deployment network.
References
Link | Resource |
---|---|
https://ics-cert.us-cert.gov/advisories/ICSMA-18-058-01 | Third Party Advisory US Government Resource |
https://ics-cert.us-cert.gov/advisories/ICSMA-18-058-01 | Third Party Advisory US Government Resource |
Configurations
Configuration 1 (hide)
AND |
|
History
21 Nov 2024, 04:08
Type | Values Removed | Values Added |
---|---|---|
References | () https://ics-cert.us-cert.gov/advisories/ICSMA-18-058-01 - Third Party Advisory, US Government Resource |
Information
Published : 2018-05-04 18:29
Updated : 2024-11-21 04:08
NVD link : CVE-2018-5446
Mitre link : CVE-2018-5446
CVE.ORG link : CVE-2018-5446
JSON object : View
Products Affected
medtronic
- 2090_carelink_programmer
- 2090_carelink_programmer_firmware