Vulnerabilities (CVE)

Filtered by vendor Medtronic Subscribe
Filtered by product Mycarelink 24950 Patient Monitor Firmware
Total 2 CVE
CVE Vendors Products Updated CVSS v2 CVSS v3
CVE-2018-10626 1 Medtronic 4 Mycarelink 24950 Patient Monitor, Mycarelink 24950 Patient Monitor Firmware, Mycarelink 24952 Patient Monitor and 1 more 2024-11-21 3.8 LOW 4.4 MEDIUM
A vulnerability was discovered in all versions of Medtronic MyCareLink 24950 and 24952 Patient Monitor. The affected product's update service does not sufficiently verify the authenticity of the data uploaded. An attacker who obtains per-product credentials from the monitor and paired implantable cardiac device information can potentially upload invalid data to the Medtronic CareLink network.
CVE-2018-10622 1 Medtronic 4 Mycarelink 24950 Patient Monitor, Mycarelink 24950 Patient Monitor Firmware, Mycarelink 24952 Patient Monitor and 1 more 2024-11-21 1.9 LOW 7.1 HIGH
A vulnerability was discovered in all versions of Medtronic MyCareLink 24950 and 24952 Patient Monitor. The affected products use per-product credentials that are stored in a recoverable format. An attacker can use these credentials for network authentication and encryption of local data at rest.