Vulnerabilities (CVE)

Filtered by vendor Bd Subscribe
Filtered by product Alaris 8015 Pc Unit
Total 2 CVE
CVE Vendors Products Updated CVSS v2 CVSS v3
CVE-2016-9355 1 Bd 1 Alaris 8015 Pc Unit 2024-11-21 2.1 LOW 5.3 MEDIUM
An issue was discovered in Becton, Dickinson and Company (BD) Alaris 8015 Point of Care (PC) unit, Version 9.5 and prior versions, and Version 9.7. An unauthorized user with physical access to an Alaris 8015 PC unit may be able to obtain unencrypted wireless network authentication credentials and other sensitive technical data by disassembling an Alaris 8015 PC unit and accessing the device's flash memory. Older software versions of the Alaris 8015 PC unit, Version 9.5 and prior versions, store wireless network authentication credentials and other sensitive technical data on the affected device's removable flash memory. Being able to remove the flash memory from the affected device reduces the risk of detection, allowing an attacker to extract stored data at the attacker's convenience.
CVE-2016-8375 1 Bd 1 Alaris 8015 Pc Unit 2024-11-21 1.9 LOW 4.9 MEDIUM
An issue was discovered in Becton, Dickinson and Company (BD) Alaris 8015 Point of Care (PC) unit, Version 9.5 and prior versions, and Version 9.7, and 8000 PC unit. An unauthorized user with physical access to an affected Alaris PC unit may be able to obtain unencrypted wireless network authentication credentials and other sensitive technical data by disassembling the PC unit and accessing the device's flash memory. The Alaris 8015 PC unit, Version 9.7, and the 8000 PC unit store wireless network authentication credentials and other sensitive technical data on internal flash memory. Accessing the internal flash memory of the affected device would require special tools to extract data and carrying out this attack at a healthcare facility would increase the likelihood of detection.