CVE-2021-33885

An Insufficient Verification of Data Authenticity vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to send the device malicious data that will be used in place of the correct data. This results in full system command access and execution because of the lack of cryptographic signatures on critical data sets.
Configurations

Configuration 1 (hide)

AND
cpe:2.3:o:bbraun:spacecom2:*:*:*:*:*:*:*:*
OR cpe:2.3:h:bbraun:infusomat_large_volume_pump_871305u:-:*:*:*:*:*:*:*
cpe:2.3:h:bbraun:spacestation_8713142u:-:*:*:*:*:*:*:*

History

21 Nov 2024, 06:09

Type Values Removed Values Added
CVSS v2 : 10.0
v3 : 9.8
v2 : 10.0
v3 : 10.0
References () https://www.bbraunusa.com/en.htm - Broken Link () https://www.bbraunusa.com/en.htm - Broken Link
References () https://www.mcafee.com/blogs/enterprise/mcafee-enterprise-atr/mcafee-enterprise-atr-uncovers-vulnerabilities-in-globally-used-b-braun-infusion-pump/ - Exploit, Third Party Advisory () https://www.mcafee.com/blogs/enterprise/mcafee-enterprise-atr/mcafee-enterprise-atr-uncovers-vulnerabilities-in-globally-used-b-braun-infusion-pump/ - Exploit, Third Party Advisory

Information

Published : 2021-08-25 12:15

Updated : 2024-11-21 06:09


NVD link : CVE-2021-33885

Mitre link : CVE-2021-33885

CVE.ORG link : CVE-2021-33885


JSON object : View

Products Affected

bbraun

  • spacecom2
  • spacestation_8713142u
  • infusomat_large_volume_pump_871305u
CWE
CWE-347

Improper Verification of Cryptographic Signature