CVE-2020-7590

A vulnerability has been identified in DCA Vantage Analyzer (All versions < V4.5 are affected by CVE-2020-7590. In addition, serial numbers < 40000 running software V4.4.0 are also affected by CVE-2020-15797). Affected devices use a hard-coded password to protect the onboard database. This could allow an attacker to read and or modify the onboard database. Successful exploitation requires direct physical access to the device.
Configurations

Configuration 1 (hide)

AND
cpe:2.3:o:siemens:dca_vantage_analyzer_firmware:*:*:*:*:*:*:*:*
cpe:2.3:h:siemens:dca_vantage_analyzer:-:*:*:*:*:*:*:*

History

21 Nov 2024, 05:37

Type Values Removed Values Added
References () https://www.siemens-healthineers.com/support-documentation/security-advisory - Vendor Advisory () https://www.siemens-healthineers.com/support-documentation/security-advisory - Vendor Advisory

Information

Published : 2020-10-13 16:15

Updated : 2024-11-21 05:37


NVD link : CVE-2020-7590

Mitre link : CVE-2020-7590

CVE.ORG link : CVE-2020-7590


JSON object : View

Products Affected

siemens

  • dca_vantage_analyzer_firmware
  • dca_vantage_analyzer
CWE
CWE-259

Use of Hard-coded Password