Medtronic MyCareLink Smart 25000 all versions contain an authentication protocol vuln where the method used to auth between MCL Smart Patient Reader and MyCareLink Smart mobile app is vulnerable to bypass. This vuln allows attacker to use other mobile device or malicious app on smartphone to auth to the patient’s Smart Reader, fools the device into thinking its communicating with the actual smart phone application when executed in range of Bluetooth.
References
Link | Resource |
---|---|
https://us-cert.cisa.gov/ics/advisories/icsma-20-345-01 | Third Party Advisory US Government Resource |
Configurations
Configuration 1 (hide)
AND |
|
History
No history.
Information
Published : 2020-12-14 20:15
Updated : 2024-02-28 18:08
NVD link : CVE-2020-25183
Mitre link : CVE-2020-25183
CVE.ORG link : CVE-2020-25183
JSON object : View
Products Affected
medtronic
- mycarelink_smart_model_25000_firmware
- mycarelink_smart_model_25000
CWE
CWE-287
Improper Authentication