An issue was discovered on Dr Trust ECG Pen 2.00.08 devices. Because the Bluetooth LE support is implemented without a requirement for pairing or security, any attacker can access the GATT server of the device and can sniff the data being broadcasted while a measurement is being done. Also, saved data can also be extracted over a Bluetooth connection. In addition, an attacker can launch a man-in-the-middle attack against data integrity.
References
Link | Resource |
---|---|
https://payatu.com/advisory/lack-of-bluetooth-le-encryption-and-access-control-in-dr-trust-ecg-or-ekg-pen | Third Party Advisory |
Configurations
Configuration 1 (hide)
AND |
|
History
No history.
Information
Published : 2020-08-26 16:15
Updated : 2024-02-28 17:47
NVD link : CVE-2020-15486
Mitre link : CVE-2020-15486
CVE.ORG link : CVE-2020-15486
JSON object : View
Products Affected
drtrust
- electrocardiogram_pen_firmware
- electrocardiogram_pen
CWE