The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) telnet Command-Line Interface, grants access to sensitive data stored on the WBM that permits temporary configuration changes to network settings of the WBM, and allows the WBM to be rebooted. Temporary configuration changes to network settings are removed upon reboot.
References
Link | Resource |
---|---|
https://www.us-cert.gov/ics/advisories/icsma-20-170-04 | Third Party Advisory US Government Resource |
https://www.us-cert.gov/ics/advisories/icsma-20-170-04 | Third Party Advisory US Government Resource |
Configurations
Configuration 1 (hide)
AND |
|
History
21 Nov 2024, 04:59
Type | Values Removed | Values Added |
---|---|---|
References | () https://www.us-cert.gov/ics/advisories/icsma-20-170-04 - Third Party Advisory, US Government Resource |
Information
Published : 2020-06-29 14:15
Updated : 2024-11-21 04:59
NVD link : CVE-2020-12041
Mitre link : CVE-2020-12041
CVE.ORG link : CVE-2020-12041
JSON object : View
Products Affected
baxter
- sigma_spectrum_infusion_system
- sigma_spectrum_infusion_system_firmware
- wireless_battery_module
CWE
CWE-732
Incorrect Permission Assignment for Critical Resource