CVE-2020-12039

Baxter Sigma Spectrum Infusion Pumps Sigma Spectrum Infusion System v's6.x model 35700BAX & Baxter Spectrum Infusion System v's8.x model 35700BAX2 contain hardcoded passwords when physically entered on the keypad provide access to biomedical menus including device settings, view calibration values, network configuration of Sigma Spectrum WBM if installed.
References
Link Resource
https://www.us-cert.gov/ics/advisories/icsma-20-170-04 Third Party Advisory US Government Resource
Configurations

Configuration 1 (hide)

AND
OR cpe:2.3:o:baxter:sigma_spectrum_infusion_system_firmware:*:*:*:*:*:*:*:*
cpe:2.3:o:baxter:sigma_spectrum_infusion_system_firmware:8.0:*:*:*:*:*:*:*
cpe:2.3:h:baxter:sigma_spectrum_infusion_system:-:*:*:*:*:*:*:*

History

No history.

Information

Published : 2020-06-29 14:15

Updated : 2024-02-28 17:47


NVD link : CVE-2020-12039

Mitre link : CVE-2020-12039

CVE.ORG link : CVE-2020-12039


JSON object : View

Products Affected

baxter

  • sigma_spectrum_infusion_system_firmware
  • sigma_spectrum_infusion_system
CWE
CWE-798

Use of Hard-coded Credentials

CWE-259

Use of Hard-coded Password