Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The PrismaFlex device contains a hard-coded service password that provides access to biomedical information, device settings, calibration settings, and network configuration. This could allow an attacker to modify device settings and calibration.
References
Link | Resource |
---|---|
https://www.us-cert.gov/ics/advisories/icsma-20-170-01 | Not Applicable |
https://us-cert.cisa.gov/ics/advisories/icsma-20-170-02 | Third Party Advisory |
https://www.us-cert.gov/ics/advisories/icsma-20-170-01 | Not Applicable |
Configurations
History
21 Nov 2024, 04:59
Type | Values Removed | Values Added |
---|---|---|
References | () https://www.us-cert.gov/ics/advisories/icsma-20-170-01 - Not Applicable |
Information
Published : 2020-06-29 14:15
Updated : 2024-11-21 04:59
NVD link : CVE-2020-12035
Mitre link : CVE-2020-12035
CVE.ORG link : CVE-2020-12035
JSON object : View
Products Affected
baxter
- prismaflex_firmware
- prismax_firmware
- prismax
- prismaflex