Baxter ExactaMix EM 2400 versions 1.10, 1.11, 1.13, 1.14 and ExactaMix EM1200 Versions 1.1, 1.2, 1.4 and 1.5 does not restrict access to the USB interface from an unauthorized user with physical access. Successful exploitation of this vulnerability may allow an attacker with physical access to the system the ability to load an unauthorized payload or unauthorized access to the hard drive by booting a live USB OS. This could impact confidentiality and integrity of the system and risk exposure of sensitive information including PHI.
References
Link | Resource |
---|---|
https://www.us-cert.gov/ics/advisories/icsma-20-170-01 | Third Party Advisory US Government Resource |
https://www.us-cert.gov/ics/advisories/icsma-20-170-01 | Third Party Advisory US Government Resource |
Configurations
Configuration 1 (hide)
AND |
|
Configuration 2 (hide)
AND |
|
History
21 Nov 2024, 04:59
Type | Values Removed | Values Added |
---|---|---|
References | () https://www.us-cert.gov/ics/advisories/icsma-20-170-01 - Third Party Advisory, US Government Resource |
Information
Published : 2020-06-29 14:15
Updated : 2024-11-21 04:59
NVD link : CVE-2020-12024
Mitre link : CVE-2020-12024
CVE.ORG link : CVE-2020-12024
JSON object : View
Products Affected
baxter
- em2400
- em1200
- em1200_firmware
- em2400_firmware
CWE