CVE-2018-19010

Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. A malformed network packet may cause the monitor to reboot. By repeatedly sending the malformed network packet, an attacker may be able to disrupt patient monitoring by causing the monitor to repeatedly reboot until it falls back to default configuration and loses network connectivity.
References
Link Resource
http://www.securityfocus.com/bid/106683 Third Party Advisory VDB Entry
https://ics-cert.us-cert.gov/advisories/ICSMA-19-022-01 Third Party Advisory US Government Resource
Configurations

Configuration 1 (hide)

AND
cpe:2.3:o:draeger:kappa_firmware:*:*:*:*:*:*:*:*
cpe:2.3:h:draeger:kappa:-:*:*:*:*:*:*:*

Configuration 2 (hide)

AND
cpe:2.3:o:draeger:infinity_explorer_c700_firmware:*:*:*:*:*:*:*:*
cpe:2.3:h:draeger:infinity_explorer_c700:-:*:*:*:*:*:*:*

Configuration 3 (hide)

AND
cpe:2.3:o:draeger:delta_xl_firmware:*:*:*:*:*:*:*:*
cpe:2.3:h:draeger:delta_xl:-:*:*:*:*:*:*:*

Configuration 4 (hide)

AND
cpe:2.3:o:draeger:infinity_delta_firmware:*:*:*:*:*:*:*:*
cpe:2.3:h:draeger:infinity_delta:-:*:*:*:*:*:*:*

History

No history.

Information

Published : 2019-01-28 22:29

Updated : 2024-02-28 16:48


NVD link : CVE-2018-19010

Mitre link : CVE-2018-19010

CVE.ORG link : CVE-2018-19010


JSON object : View

Products Affected

draeger

  • delta_xl_firmware
  • kappa_firmware
  • infinity_delta
  • delta_xl
  • infinity_delta_firmware
  • infinity_explorer_c700_firmware
  • infinity_explorer_c700
  • kappa
CWE
CWE-20

Improper Input Validation