An issue was discovered in Roche Accu-Chek Inform II Instrument before 03.06.00 (Serial number below 14000) and 04.x before 04.03.00 (Serial Number above 14000), CoaguChek Pro II before 04.03.00, CoaguChek XS Plus before 03.01.06, CoaguChek XS Pro before 03.01.06, cobas h 232 before 03.01.03 (Serial number below KQ0400000 or KS0400000), and cobas h 232 before 04.00.04 (Serial number above KQ0400000 or KS0400000). A vulnerability in the software update mechanism allows authenticated attackers in the adjacent network to overwrite arbitrary files on the system through a crafted update package.
References
Link | Resource |
---|---|
http://www.securityfocus.com/bid/105843 | Third Party Advisory VDB Entry |
https://ics-cert.us-cert.gov/advisories/ICSMA-18-310-01 | Mitigation Third Party Advisory US Government Resource |
Configurations
Configuration 1 (hide)
AND |
|
Configuration 2 (hide)
AND |
|
Configuration 3 (hide)
AND |
|
Configuration 4 (hide)
AND |
|
Configuration 5 (hide)
AND |
|
History
No history.
Information
Published : 2018-11-20 19:29
Updated : 2024-02-28 16:48
NVD link : CVE-2018-18565
Mitre link : CVE-2018-18565
CVE.ORG link : CVE-2018-18565
JSON object : View
Products Affected
roche
- accu-chek_inform_ii
- accu-chek_inform_ii_firmware
- coaguchek_xs_plus
- coaguchek_xs_pro
- coaguchek_xs_plus_firmware
- coaguchek_xs_pro_firmware
- coaguchek_pro_ii
- cobas_h_232_firmware
- cobas_h_232
- coaguchek_pro_ii_firmware
CWE
CWE-434
Unrestricted Upload of File with Dangerous Type