CVE-2014-5433

An unauthenticated remote attacker may be able to execute commands to view wireless account credentials that are stored in cleartext on Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16, which may allow an attacker to gain access the host network. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.
References
Link Resource
https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01 Mitigation Third Party Advisory US Government Resource
https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01 Mitigation Third Party Advisory US Government Resource
Configurations

Configuration 1 (hide)

AND
cpe:2.3:o:baxter:sigma_spectrum_infusion_system_firmware:6.05:*:*:*:*:*:*:*
OR cpe:2.3:h:baxter:sigma_spectrum_infusion_system:-:*:*:*:*:*:*:*
cpe:2.3:h:baxter:wireless_battery_module:16:*:*:*:*:*:*:*

History

21 Nov 2024, 02:12

Type Values Removed Values Added
References () https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01 - Mitigation, US Government Resource, Third Party Advisory () https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01 - Mitigation, Third Party Advisory, US Government Resource

Information

Published : 2019-03-26 16:29

Updated : 2024-11-21 02:12


NVD link : CVE-2014-5433

Mitre link : CVE-2014-5433

CVE.ORG link : CVE-2014-5433


JSON object : View

Products Affected

baxter

  • sigma_spectrum_infusion_system
  • sigma_spectrum_infusion_system_firmware
  • wireless_battery_module
CWE
CWE-312

Cleartext Storage of Sensitive Information

CWE-255

Credentials Management Errors